Inappropriate Positioning of Separator Gel in Blood Collection Tube: A Case Report

Main Article Content

Wan Norlina Wan Azman
Noorazliyana Shafii
Siong Hu Wong
Aniza Mohd Jelani

Abstract

Blood collection tubes containing separator gel are widely used in many clinical laboratories because of easy to use, convenience and several other advantages. However, they are not void of limitation. We describe an unusual case with a floating separator gel in the primary blood collection tube. The blood sample was collected into a BD Vacutainer SST II Advance tube, from a 51 years old man admitted for community acquired pneumonia and had his- tory of hyperproteinaemia. Inappropriate positioning of the separator gel was observed after centrifugation, whereby the separator gel floated above the serum. Detection of abnormal separator gel floatation pattern in this patient’s sample before analysis had prevented potential technical problem and production of erroneous result caused by aspiration of separator gel. The limitation of blood collection tube with separator gel especially in patients with hy- perproteinaemia should be made aware among laboratory personnel and physicians.

Downloads

Download data is not yet available.

Article Details

How to Cite
Wan Azman, W. N., Shafii, N., Wong, S. H., & Mohd Jelani, A. (2021). Inappropriate Positioning of Separator Gel in Blood Collection Tube: A Case Report. Malaysian Journal of Medicine and Health Sciences, 17(3), 308–310. Retrieved from http://mjmhsojs.upm.edu.my/index.php/mjmhs/article/view/441
Section
Case Report

References

Bowen RA, Hortin GL, Csako G, Otañez OH, Remaley AT. Impact of blood collection devices on clinical chemistry assays. Clinical Biochem. 2010;43(1-2):4-25.

van den Ouweland JM, Church S. High total protein impairs appropriate gel barrier formation in BD Vacutainer blood collection tubes. Clin Chem. 2007;53(2):364-5.

Haklar G, Sirikci O, Baykan O, Ramazan DC, Gerin F. Abnormal gel flotation in a patient with apperant pneumonia diagnosis: a case report. Biochemia Medica: 2014;24(1):180-2.

Fatás M, Franquelo P, Franquelo R. Anomalous flotation of separator gel: density or viscosity? Clin Chem. 2008;54(4):771-2.

Fortgens P, Pillay TS. Pseudohyponatremia revisited: a modern-day pitfall. Arch Pathol Lab Med. 2011;135(4):516-9.